OTHER DETECTOR EXPERIENCE Robert Schofield is located at LHO and plans to continue gaining general experience with the detector - invaluable in understanding the detector operation. So far, in addition to experience in the environmental monitoring areas mentioned above he has: 1) Helped install seismic isolation systems in HAM-10, BSC-8, BSC-5 and BSC-4. 2) Measured the changing tilt of the concrete slab as WHAM - 7 was pumped down. The difference in tilt was about 5 microradians. This tilt will affect laser alignment 3) Helped test chamber shakers and accelerometers. 4) Helped install tilt meters and seismometers. 5) Helped install and adjust the recycling mirror in HAM-9. 6) Designed and tested an extension to monitor dust in vacuum chambers during installations. 7) Written the device driver code, EPICS interface and control room screens for the dust monitoring system.